A 40 µF capacitor in a defibrillator is charged to 3000 V. The energy stored in the…
Physics · JEE Main · NTA Exams — Electrostatics
A 40 µF capacitor in a defibrillator is charged to 3000 V. The energy stored in the capacitor is sent through the patient during a pulse of duration 2 ms. The power delivered to the patient is :–
- 45 kW
- 90 kW
- 180 kW
- 360 kW
Answer
(B) 90 kW
Sign up free on Edukali to view the step-by-step worked solution and practice thousands of similar questions.